DETERMINATION OF
UNCERTAINTY OF SULFONATED GRAFTED LOW DENSITY POLYETHYLENE DOSIMETER FOR
ASSURING THE QUALITY OF GAMMA MEASUREMENTS DIAB, H.M.
Radiation Dosimetry Laboratory, National Institute of Standards, Giza,
Egypt
Key words:
Sulfonated grafted polymers, Thermoluminescence,
إيجاد اللايقين فى القياس الإشعاعى للبوليمرات المطعمة المسلفنة
حسن محمد دياب خلاصـة
تم فى هذا البحث تشعيع مقياس من مقاييس جرعات الوميض الحرارى (البولى ايثيلين
منخفض الكثافة والمطعم بحمض الاكريليك والاكريلاميد) بجرعة جاما العلاجية من
مصدر الكوبلت60. ABSTRACT In this study, the type LDPE-g-p (AAm/AAc) of TLD dosimeter film was irradiated by 60Co therapeutic gamma ray dose. The study deals with the case of sulfonated grafted film which has good properties such as thermal stability and good dosimetric properties. The various sources of uncertainty for this type of TLD under study were analyzed. The uncertainty budget tables for radiation measurements were declared. For the used procedure, these uncertainties were multiplied by the coverage factor (K = 2) to obtain the expanded uncertainty at 95% confidence level. The combined uncertainty didn’t exceed 6.3%. The expanded uncertainty at 95% confidence level should be added to the value of measurements to obtain the accurate dose.
|
|
|