DETERMINATION OF THE EQUILIBRIUM FACTOR
BETWEEN RADON AND ITS PROGENY BY ACTIVE
AND PASSIVE TECHNIQUES
*

ARAFA, W.(1), EL-SERSY, A.(2), EL-SAMMAN, H.(3) and BADRAN, H.(4) 

(1) Physics Dept., Faculty of Women, Ain Shams University, Cairo, Egypt,
(2) Radiation Dosimetry Dept., National Institute of Standards,  Giza, Egypt,
(3) Physics Dept., Faculty of Science , Menoufya University, Shebin
El-Koom, Egypt and (4) Physics Dept., Faculty of Education,
Ain Shams University, Cairo, Egypt.

Key words: Radon, Equilibrium factor, Equivalent dose, Surface
                  barrier detector, CR-39 detector, Charcoal canisters.

تعيين معامل الاتزان بين غاز الرادون ووليداته باستخدام
طرق القياس المباشرة
والغير مباشرة

وفاء عرفة و أحمد السرسى و حسين السمان و هبة بدران

خلاصـة

تم تحديث غرفة قياس غاز الرادون (0.241 م3) بهدف استخدامها لتعيين معامل الاتزان بين غاز الرادون ووليداته  لكواشف الكربون المنشط والبوليمر CR-39 وكاشف السليكون الحائل السطحى.

أوضحت النتائج أن قيمة معامل الاتزان التى تم الحصول عليها بإستخدام كاشف السليكون تتفق بدرجة كبيرة مع تلك القيم التى تم الحصول عليها باستخدام كاشف CR-39 كما أوضحت النتائج أيضاً أن قيمة معامل الاتزان تتأثر بالتغيرات المناخية المحيطة. وقد وجد ان قيمة معامل الاتزان التى تم الحصول عليها باستخدام كاشف الكربون المنشط لا تتأثر كثيراً بالمتغيرات المناخية المحيطة بالكاشف و يجب أن يعدل بمقدار 0.482 ليتوافق مع قيمة المعامل فى حالة كاشف السليكون. تم أيضاً حساب الجرعة الإشعاعية المكافئة لغاز الرادون باستخدام الكواشف الثلاثة  وكان هناك توافق فى النتائج  التى تم الحصول عليها.

ABSTRACT

A 0.241 cubic meter radon chamber was updated and used to determine the equilibrium factor (F) between radon and its progeny.  The value of "F" was determined using an active technique for surface barrier (SB) detector and also a passive technique for CR-39 and activated charcoal canister (ACC) detectors. The effect of environmental conditions on "F" value was also investigated. A comparison between different values of "F" showed that the values obtained by CR-39 and SB detectors were in coherent agreement and were sensitive to the variations of environmental conditions. However, "F" values obtained by ACC detector were higher than those obtained by SB and CR-39 detectors by a constant factor. The radon equivalent dose calculated by different techniques was found to be in agreement.


 

Accepted May 2009.

 

 

 

 

اهداف المركز
أ
هداف المركز

الهيكل التنظيمى
الهيكل التنظيمى


الخدمات
الخدمات


المجلة
المجلة


الدورات التدريبية
الدورات التدريبية


منح الزمالة
منح الزمالة


العنوان: 9 شارع ملاعب الجامعة   الدقى جيزة جمهورية مصر العربية          رقم بريدى:12311
تليفون: 3370588 - 3370569                                    تليفاكس: 3371082
البريد الإلكتروني : merrcac@yahoo.com